Cineloop Check; Preparation; Adjust The Cineloop Controls - GE LOGIQ V2 Basic Service Manual

Hide thumbs Also See for LOGIQ V2:
Table of Contents

Advertisement

General Procedures and Functional Checks

Cineloop Check

Introduction

Preparation

Adjust the Cineloop controls

4-38
A cineloop is a sequence of images recorded over a certain time
frame. When using ECG the time frame can be adjusted to
cover one or more heart cycles. When frozen, the System
automatically displays the cineloop boundary markers on either
side of the last detected heart cycle.
Connect one of the probes to the scanner.
See
'Probe/Connectors Check' on page 4-37
connecting the probes
For available probes, see
page
9-9:
Turn ON the scanner. The 2D Mode is displayed (default
mode).
Press Freeze.
The left and right markers are displayed on either side of the
last detected heart cycle on the ECG trace.
Press Freeze.
The selected heart beat is played back.
Press Freeze to freeze the cineloop.
Use the trackball to scroll through the acquisition and find
the sequence of interest.
Adjust Cycle select to move from heart beat to heart beat
and select the heart cycle of interest.
Adjust Num cycles to increase or decrease the number of
heart beats to be played back.
Adjust Left marker and Right marker to trim or expand the
cineloop boundaries.
LOGIQ V2/LOGIQ V1 – Basic Service Manual
for info about
'Probe and Accessories' on
5610739-100 English Rev.10