Replacement Procedures
Software Upgrade Procedure (For R1.1.x) (continued)
7-32
8. The system will display upgrading result and shutdown
automatically within 30 seconds.Or click the Shutdown in
the screen to shut down the system immediately.
Figure 7-20. System upgrade complete
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10