System Setup
Setup warnings (continued)
NOTE:
DANGER Equipment damage possibility. Turning the system on without
acclimation after arriving at site may cause the system to be
damaged.
CAUTION
50
Degree C
122
Degree F
4
hours
3-4
1. There are no operator serviceable components. To prevent
shock, do not remove any covers or panels. Should
problems or malfunctions occur, unplug the power cord.
Only qualified service personnel should carry out servicing.
For information regarding packing labels, refer to LABELS
ON PACKAGE.
2. After being transported, the unit may be very cold or hot. If
this is the case, allow the unit to acclimate before you turn it
on. It requires one hour for each 2.5°C increment it's
temperature is below 3°C or above 40°C.
If the Ultrasound system is very cold or hot, do not turn on its power
until it has had a chance to acclimate to its operating environment.
The following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
Table 3-1:
System Acclimation Time Chart