Warnings And Important Information; Warnings - GE LOGIQ V2 Basic Service Manual

Hide thumbs Also See for LOGIQ V2:
Table of Contents

Advertisement

Warnings and important information

Warnings

WARNING
WARNING
WARNING
NOTE:
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
Energy Control and Power Lockout for LOGIQ V2/LOGIQ V1.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Ultrasound System components may be energized.
Because of the limited access to cabinets and equipment in the
field, placing people in awkward positions, GE has limited the
lifting weight for one person in the field to 16 KG (35 LBS).
Anything over 16 KG (35 LBS) requires 2 people.
DO NOT touch any boards with integrated circuits prior to
taking the necessary ESD precautions.
Always connect yourself, via an arm-wrist strap, to the advised
ESD connection point located on the rear of the Ultrasound
system (near the power connector).
Follow general guidelines for handling of electrostatic sensitive
equipment.
Use an ESD compatible work space or the ESD-kit during parts
replacement.
Warnings and important information
7-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq v156612445661242

Table of Contents