GE LOGIQ V2 Basic Service Manual page 303

Hide thumbs Also See for LOGIQ V2:
Table of Contents

Advertisement

Data Sheet for enclosure Source Leakage Current
Table 10-16:
Tester
Unit
Polarity
Power
Switch
Enter Name of tested peripheral here:
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
The test passes when all readings measure less than the value
shown in Table 10-12 on page 10-22. Record all data on the PM
Inspection Certificate.
Typical Data Sheet for enclosure Source Leakage Current
Tester
Neutral
or
Test 1
Ground
Speaker
Switch
Cover
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Electrical safety tests
Test 2 Real
Panel Metal
Optional
Parts
Test 3
Optional
Test 4
10-27