General Procedures and Functional Checks

Adjust the PW Doppler Mode controls

NOTE:
NOTE:
4-34
Adjust the Active mode gain to set the gain in the spectral
Doppler area.
Adjust Low velocity reject to reduce unwanted low velocity
blood flow and tissue movement.
In PW mode, adjust Sample volume to low setting for better
resolution, or higher setting to more easily locate the
disturbed flows.
Adjust the Compress setting to balance the effect of
stronger and weaker echoes and obtain the desired intensity
display.
Adjust Frequency to optimize flow display. Higher setting
will improve resolution and the lower setting will increase the
depth penetration.
Adjust Frame rate to a higher setting to improve motion
detection, or to a lower setting to improve resolution.
Frequency and Frame rate settings may affect the Low
Velocity Reject.
The Doppler Power setting affects only Doppler operating
modes.
Adjust the following settings to further optimize the display
of the image.
Use the Horizontal sweep to optimize the sweep speed.
To view signal detail, adjust Scale to enlarge the vertical
spectral Doppler trace.
Use Invert to reverse the vertical component of the spectral
Doppler area of the display.
Use Angle correction to steer the ultrasound beam to the
blood flow to be measured.
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10