GE H
EALTHCARE
D
5344303-100, R
IRECTION
5) Remove the two flexi cable from TX board as shown below in
6) Remove the Screws fixed in Tx Pwa & seperate the RX PWA from TX PWA as shown in
on page 8-70
8-7-0-6
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-70
PROPRIETARY TO GE
3
EVISION
Figure 8-61 Removing TX conn board cable & Flexi cable
Figure 8-62 Tx-RX combined
Section 8-7 - RX Board (FRU P/N: 5315027)
VIVID P3 S
Figure 8-61 on page 8-70
Tx conn board cable
Flexi cable
M
ERVICE
ANUAL
Figure 8-62