8-13-0-2 Initiate software reload from the Recovery Console
Ensure that the VIVID P3 is powered down.
1.) Depress the On/Off button on the Operator Panel. The VIVID P3 starts.
2.) Press the Esc button, on the alphanumeric keyboard, multiple times until the GRUB menu is
displayed on the screen.
3.) On the Grub menu, select: Recovery Console.
The GE Healthcare Recovery Console is displayed.
Enter the Pass word
4.) Call the Online Center (OLC).
Tell OLC that you need a "Challenge Code", and on request, you must read the Challenge string on
the screen. OLC will return a password that you need to type in as the Response.
5.) Enter the Response (password) you got from OLC.If, accepted, the System Software Loading from
Repository, starts.
6.) After the Software loading is complete Restore the Backed up Database.
PROPRIETARY TO GE
3
EVISION
Figure 8-80 GRUB menu
Chapter 8 Replacement Procedures
VIVID P3 S
M
ERVICE
ANUAL
8-91