GE Vivid P3 Service Manual page 117

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
4-3-10-2
Doppler Mode OP Panel Controls
Step
1
Press CW Mode key
2
Adjust Gain
3
Display M/D-Mode Cursor
4
B-Pause
PROPRIETARY TO GE
3
EVISION
Table 4-11 Doppler Mode OP Panel Controls
Task
Chapter 4 Functional Checks
Expected Result(s)
CW Mode Starts
Amplifies the overall strength of the echoes
processed in the Color Flow window. Turn the
Gain dial (PW Mode key) to the left/right to
increase/decrease Gain.
Displays the M/D-Mode cursor on the B-Mode
image. Press Cursor and Trackball to position
sample volume graphic. Click SV gate to
adjust sample volume gate size.
Toggle between simultaneous and update
presentation while viewing Spectral Doppler.
Press B Pause to toggle between
simultaneous and update.
V
P3 S
M
IVID
ERVICE
ANUAL
4-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

53503755350378535037953503805350381

Table of Contents