GE Vivid P3 Service Manual page 264

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
8-7-1
Flexi Cable (MST to TX) Assy (FRU No.5315109)
This is a description on how to remove and replace the TX Board or Flex cable.
8-7-1-1
Tools
Common Phillips screwdrivers
8-7-1-2
FRU BOM. This FRU Kit P/N:5315109 consists of following parts.
PART NO
5268515
5314200
8-7-1-3
Needed Manpower
1 person, 30 minutes
8-7-1-4
Preparations
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-7-1-5
Removal Procedure
1) Remove the connector board assembly. Refer
2) Loosen the Five screws (1,2,3,4,5,) of TX board from the system. Refer
3) Remove the Tx board from the system. Refer
8-72
PROPRIETARY TO GE
3
EVISION
DESCRIPTION
Flexi Cable (MST to TX)
Fastener-M2X3 PAN PHILIP M/C SCREW (P1/P2)
1
2
3
Figure 8-63 TX board Assembly
Section 8-7 - RX Board (FRU P/N: 5315027)
section 8-6-2 on page 63
Figure 8-59 on page 8-67
4
5
VIVID P3 S
M
ERVICE
ANUAL
Figure 8-59 on page 8-67