10-7-6-3
Dale 600 Meter Procedure (cont'd)
5.) Set the meter's "FUNCTION" switch to LEAD TO GROUND position to measure the patient lead to
ground leakage current.
6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the
power condition combinations.
Table 10-18
PROPRIETARY TO GE
3
EVISION
Figure 10-7 ECG Leakage Current Test
Testing Power Conditions
ECG Power
Meter's Polarity Switch
ON
ON
ON
ON
OFF
OFF
OFF
OFF
Chapter 10 Care & Maintenance
Meter's Neutral Switch
NORM
NORM
REVERSE
REVERSE
NORM
NORM
REVERSE
REVERSE
VIVID P3 S
M
ERVICE
ANUAL
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
10-21