Transducer Connection - GE Vivid P3 Service Manual

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
3-5-3

Transducer Connection

1.) Connect a transducer to the upper transducer receptancle as follows:
A.) Ensure that the transducer twist lock lever points towards the 9 o'clock position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating
surface.
C.) Twist the transducer lock lever vertically to lock it in place. Twist the lever to the horizontal position
to disconnect the transducer.
NOTE:
Please ensure that the probe latch is in unlocked position before you connect the probe to the system.
Fig (A) Connecting the Probe
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
PROPRIETARY TO GE
3
EVISION
Figure 3-8 Connect the probe
Chapter 3 Installation
Fig (B) Locking the Probe Lever
VIVID P3 S
M
ERVICE
ANUAL
3-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

53503755350378535037953503805350381

Table of Contents