10-7-2
GEMS Leakage Current Limits
The following limits are summarized for NFPA 99 (For USA), IEC 60601-1 Medical Equipment Safety
Standards, and IEC 62353 Medical Electrical Equipment — Recurrent test and test after repair of
medical electrical equipment. Measurement limits per IEC 60601-1 Medical Equipment Safety
Standards, Table IV.
Table 10-13 Chassis Leakage Current Limits—Accessible Metal Surfaces
Normal Condition
Table 10-14 Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating) Surface
All Countries
Table 10-15 Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG
Normal Condition
*
NOTE:
*Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table IV
**Most meters (like Dale 600/601) measure a composite of the AC and DC leakage current.
PROPRIETARY TO GE
3
EVISION
0.1 mA
and Cavity Probes
Normal Condition
0.1 mA
AC
0.01 mA
DC**
Connections
0.01 mA
Chapter 10 Care & Maintenance
Open Ground
0.5 mA
Open Ground
0.5 mA
0.05 mA
Open Ground
0.05 mA
VIVID P3 S
M
ERVICE
ANUAL
Open Neutral
0.5 mA
Open Neutral
0.5 mA
0.05 mA
Open Neutral
0.05 mA
10-13