4-3-9-2
Doppler Mode OP Panel Controls
Step
1
Press PW Mode key
2
Adjust Gain
3
Display M/D-Mode Cursor
4
B-Pause
4-20
PROPRIETARY TO GE
3
EVISION
Table 4-9 Doppler Mode OP Panel Controls
Task
Section 4-3 - General Procedure
Expected Result(s)
PW Mode Starts
Amplifies the overall strength of the echoes
processed in the Color Flow window. Turn the
Gain dial (PW Mode key) to the left/right to
increase/decrease Gain.
Displays the M/D-Mode cursor on the B-Mode
image. Press Cursor and Trackball to position
sample volume graphic. Click SV gate to
adjust sample volume gate size.
Toggle between simultaneous and update
presentation while viewing Spectral Doppler.
Press B Pause to toggle between
simultaneous and update.