GE Vivid P3 Service Manual page 327

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
10-7-9-2
Test Equipment (cont'd)
.
Table 10-23 Test Equipment and Accessory Description
Dale Part number
Vendor MODEL
600/100
46-285647P2
Vendor MODEL
600/102
46-285647P4
Vendor MODEL
600/200
46-285647P6
PROPRIETARY TO GE
3
EVISION
Accessory Name
CHASSIS CABLE
CHASSIS GROUND PROBE
ISO/EXTERNAL LEAKAGE
CABLE
Chapter 10 Care & Maintenance
Picture
Black coil with extended length
and black grips
Black coil cord with needle probe
for testing receptacles and for tight
spaces.
Black coiled cord with red grips.
VIVID P3 S
M
ERVICE
ANUAL
Description
Used on DALE601/601E to measure
earth resistance and enclosure
leakage current. Also used as
reference lead for external
measurement.
Also referred to as "Saline Probe" or
"Saline Bath Ground Cable".
Measures earth resistance and
enclosure leakage current. Also used
for grounding saline baths for
isolation testing of probes.
Used on DALE601/601E. This probe
may be substituted for the 600/100
Chassis Cable, and used as a probe
instead of a clamp.
Standard auxiliary cable for external
measurements of leakage current
and voltage gradient between two
surfaces.
Used on DALE601/601E to measure:
• Point-to-Point Leakage Current
• Probe and Transducer Isolation
Current
May only be connected to the female
connector of the Analyzer, labeled
EXTERNAL
10-25