GE Vivid P3 Service Manual page 121

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
4-3-12
ECG Checks
4-3-12-1
Introduction
The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be viewed on the
Screen.
4-3-12-2
ECG Check
Step
Connect the ECG harness to the connector on the
1
front of the system
Connect the three leads to a ECG simulator,
or
2
Fasten the three ECG Pads to your body and connect
the three leads to respective ECG Pad
PROPRIETARY TO GE
3
EVISION
Task
Chapter 4 Functional Checks
Table 4-13 ECG Check
The unit displays a straight curve along the bottom
edge of the image sector on the screen.
When connecting, the signal on the screen will be
noisy. When the connection is completed, a typical
clean ECG signal is displayed.
V
P3 S
IVID
ERVICE
Expected Result(s)
M
ANUAL
4-29

Advertisement