GE H
EALTHCARE
D
5344303-100, R
IRECTION
Section 8-8 Power Block
8-8-1
Transformer Assembly (FRU No. 5196448) Cable Guard Clamp (FRU
No:5219747)
This is a description on how to remove and replace the Transformer Assy.
8-8-1-1
Tools
•
Common Phillips screwdrivers
8-8-1-2
Needed Manpower
•
1 persons, 15 minutes
8-8-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
NOTE:
For 110V System Configuration use 110V wired Plug (with 3 Amps Fuse).
NOTE:
For 230V System Configuration use 230V Wired plug (With 2 Amps Fuse).
NOTE:
To replace Cable Guard Clamp Donot remove the transformer,remove power cord and press
the cable clamp from both the sides the clamp will come out side.
NOTE:
Before Replacing the transformer remove the Labels stuck on the old transformer and stuck on
new transformer.
8-8-1-4
Removal Procedure
1) Remove rear cover. Refer
2) Unscrew two screws (a,b) as marked in
3) Remove GND cable (which ever is connected near to "a" ) Refer
GND
cable
PROPRIETARY TO GE
3
EVISION
section 8-4-3 on page 30
Figure 8-66 Removing Transformer Assembly
Chapter 8 Replacement Procedures
Figure 8-66 on page 8-75
Figure 8-66 on page 8-75
a
b
VIVID P3 S
M
ERVICE
ANUAL
8-75