GE Vivid P3 Service Manual page 330

Hide thumbs Also See for Vivid P3:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5344303-100, R
IRECTION
10-7-9-5
Data Sheet for Transducer Source Leakage Current
The test passes when all readings measure less than the values shown in
on the PM Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
CAUTION
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
.
Table 10-24 Typical Data Sheet For Transducer Source Leakage Current
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
10-28
PROPRIETARY TO GE
3
EVISION
Tester Power Polarity
Switch
NORM
NORM
REV
REV
NORM
NORM
REV
REV
Section 10-7 - Electrical Safety Tests
Transducer Tested:
Tester GROUND or
NUETRAL Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
VIVID P3 S
M
ERVICE
ANUAL
Table
10-13. Record all data
Measurement