CLMD16 User's Manual
__
Data Damping Period
You can configure a damping parameter to smooth the tank level or pressure/vacuum readings or
make them more responsive. The data damping is configurable between 0.2 – 25.0 seconds. The
default data damping for a channel is 15.0 seconds.
Realtime Resistance
(Resistive Channels Only)
This read-only field shows the resistance currently being measured on the applicable resistive input
signal.
Realtime Current
(Current Loop Channel Only)
This read-only field shows the current being measured on the current loop input signal.
Tank Levels Calibration
Pressing this button opens the Tank Calibration dialog, shown in Figure 23.
There are two methods of calibration: Manual Table and Step Fill Table
Manual Table
Using the manual table method, you enter each entry of the measured parameter (depth) and the
level of fluid in the tank. The table may have as few as 2 entries (for a tank with rectangular cross-
section) or as many as 16 entries (for a tank with a complex cross-section).
55