Unitek Miyachi LW250 Operation, Maintenance And Service Manual page 89

Yag laser
Table of Contents

Advertisement

Code
Indication on
No.
Control Panel Screen
14 -
Branch Shutter Trouble for
17
Nos. 1 through 4
18
No Light
19
Over Limit of Laser Power
20
Under Limit of Laser Power
23
Phase Trouble
24
Memory Trouble
25
Set Error Over Limit of Max
Power
27
Pump Overheating
28-29
Bank A/B Alarm
30
No Current
31
IGBR Trouble
990-605 REV B: May, 1997
High
Voltage
OFF
OFF
OFF
OFF
OFF
OFF
LW250 PULSED Nd: YAG LASER
CHAPTER 8: SERVICE
Abnormal
Output
ON
A laser Start signal has been input
while the branch shutter or time
branch unit is moving.
The laser output is below 0.3 J.
Reconfirm the upper limit of the set
value. Change the upper limit.
Verify the lower limit setting or the
flash lamp. Change the setting or
replace the flash lamp.
ON
Check the three-phase input power
wiring.
ON
Memory backup lithium battery
charge is low. Replace the battery.
ON
The input power setting is beyond
the maximum capacity (voltage,
pulse width, repetition rate). The
setting will be restored to the values
before the change.
ON
Cooler pump is overloaded. Check
Deionized water path for blockage.
ON
Frequent condition change activates
the thermostat inside the capacitor
bank. Increase the interval of the
condition changes. The cooling fan
motor's air filter is clogged in the rear
of the laser. This caused the
overheating of the power supply.
Clean the filter.
ON
Laser is turned on but no current is
flowing in the flash lamp. Replace
the defective lamp.
ON
Current is flowing to the flash lamp,
but the laser is turned off. Probable
defective power supply. Contact
Unitek Miyachi.
Remedy
8-16