APPENDIX B: EXTERNAL COMMUNICATIONS INTERFACE SPECIFICATION
1. Data Setting (continued)
(7)
If the data number is set to [99], write the data in one line, as follows:
(Data No. 1), (Data No. 2), Data No.3), ---, (Last Data No.)
Enter a comma (,) between data, excluding the monitor values (DI water temperature,
total count, GOOD count, energy monitor value, peak power monitor value).
(8)
Valid only for external communication control. Return (NAK) for other control modes.
2.

Reading the data

S
C
Host
T
H
Computer
X
1
Side
Laser Side
(1)
CH1 and CH0 are network numbers, where CH1 = number of tens and CH0 = number of
units.
(2)
SH1 and SH0 are schedule numbers, where SH1 = number of tens and SH0 = number of
units.
(3)
DT1 and DT0 are data numbers, where DT1 = number of tens and DT0 = number of
units.
(4)
For data numbers, refer to Section 6, Contents of Set Data and Monitor Data.
(5)
For System Setup, set the schedule number to [99] (refer to Section 6(1)).
(6)
If the data number is set to [99], read the data in one line, as follows:
(Data No. 1), (Data No. 2), (Data No. 3), ---, (Last data No.)
Enter a comma (,) between data.
(7)
Return (NAK) when the schedule number or data number is out of range.
990-605 REV B: May, 1997
C
S
S
D
D
E
B
R
H
H
H
T
T
T
C
O
1
O
1
O
X
C
LW250 PULSED Nd: YAG LASER
Code: R
S
E
B
T
d a t a
T
C
X
X
C
C
C
N
or
H
H
A
1
0
K
B-10