Error List - Unitek Miyachi LW250 Operation, Maintenance And Service Manual

Yag laser
Table of Contents

Advertisement

APPENDIX B: EXTERNAL COMMUNICATIONS INTERFACE SPECIFICATION
7.

Error List

No.
00
Normal
01
Power supply cover open
02
Laser head cover open
03
Cooling water temperature high
04
Cooling water temperature low
05
Cooling water flow rate low
06
Cooling water level low
07
Conductivity caution
08
Conductivity abnormal
09
Preliminary discharge abnormal
10
Emergency stop
11
Charge abnornal (charge/discharge
overtime)
12
Duty cycle too high (input power)
13
Main shutter error
14
Branch Shutter No. 1 error
15
Branch Shutter No. 2 error
16
Branch Shutter No. 3 error
Branch Shutter No. 4 error
17
18
Branch Shutter No. 5 error
19
Branch Shutter No. 6 error
990-605 REV B: May, 1997
Error
LW250 PULSED Nd: YAG LASER
No.
20
No light
21
Laser power upper limit error
22
Laser power lower limit error
23
Not used
24
Not used
25
Memory error
26
Set up value too high
27
Fiber 1 error
28
Fiber 2 error
29
Not used
30
Bank alarm
31
Not used
32
Not used
33
Duty cycle over (charge not ready)
34
Initialize
35
Not used
36
Not used
37
Branch unit cover open
Error
B-18