Cleaning the OWA probe tip
In order to secure correct measurements, it is important to make sure that the probe system is always kept
clean. Therefore, please follow the below illustrated instructions on how to remove e.g. cerumen from the
small acoustic channels of probe tip.
D-0113323-F – 2020/11
Sera™ - Instructions for Use - US
Use the Thornton Bridge & Implant Floss or the Thornton 3 in 1 Floss
for cleaning.
Discard used floss after use.
Never clean the probe tip while the tip is still attached to the probe
body.
Unscrew the probe cap by turning it in a counterclockwise direction
and remove the probe tip from the probe body.
Thread the cleaning floss into one of the channels from the base of
the probe tip.
Pull the cleaning floss completely through the probe tip channel.
Repeat for each channel if required.
Page 65