Charging and powering Sera

3.12.1 The battery

The instrument is powered by a rechargeable Lithium-ion (Li-ion) battery. The battery is charged by placing
the instrument in the AC powered cradle.
The battery can also be charged by:
Connecting the instrument to a powered PC using the supplied USB cable. Charging this way will be
slower than charging with the cradle.
Connecting the device to mains using the power supply from the cradle directly plugged into the micro
USB port of the Sera
Systems – shall comply with the safety requirements stated the general standard IEC 60601-1, edition 3,
clause 16. Any equipment not complying with the leakage current requirements in IEC 60601-1 shall be kept
outside the patient environment i.e. at least 1.5 m from the patient support or shall be supplied via a
separation transformer to reduce the leakage currents. Any person who connects external equipment to
signal input, signal output or other connectors has formed a Medical Electrical System and is therefore
responsible for the system to comply with the requirements. If in doubt, contact qualified medical technician
or your local representative. If the instrument is connected to a PC (IT equipment forming a system), ensure
not to touch the patient while operating the PC.
If the instrument is connected to a PC (IT equipment forming a system) assembly and modifications shall be
evaluated by qualified medical technician according to safety regulations in IEC 60601.
Note! The USB connection on the instrument has insulation protections to the patient connections.
The consequence of not following this precaution could be a too high leakage current to the patient.
IMPORTANT NOTE: The internal battery should only be replaced by an authorized service representative.
Damage to the electronics resulting from an attempt to change the battery by someone other than an
authorized representative will not qualify for repair under the product warranty.

Battery lifetime and charging time

Please observe the following precautions:
Please note:
The battery can only be maintained by service personnel.
When fully charged, the Sera
default power save and power off is enabled.
D-0113323-F – 2020/11
Sera™ - Instructions for Use - US
TM
.
TM
This equipment is intended to be connected to other equipment thus forming a
Medical Electrical System. External equipment intended for connection to signal
input, signal output or other connectors shall comply with the relevant product
standard e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for medical
electrical equipment. In addition, all such combinations – Medical Electrical
Keep the battery fully charged
Do not place the battery in a fire or apply heat to the battery
Do not damage the battery or use a damaged battery
Do not expose the battery to water
Do not short circuit the battery or reverse the polarity
The battery automatically starts charging when the Sera
the AC-powered cradle or when connected via USB to the PC
Use only the charger provided with the Sera
The Sera
will shut down automatically when the battery charge is too low to
TM
support continued operation.
battery will last for a full day of testing under normal test conditions when the
TM
is placed properly into
TM
TM
Page 19