Start Calibration Menu

Select Gas to be Calibrated
Recovery Duration
Calibration Gas Values
All gases selected under Sensor Configuration (O
require calibration but do not have to be calibrated at the same time. However,
calibration will always occur in the same sequence.
1. O
2. Zero
3. Combustibles Span
4. Methane Span
For example if the sensor is configured for oxygen, combustibles and methane
and you are skipping the O
1. Zero Calibration
2. Combustible Span Calibration
3. Methane Span Calibration
PN 9000-165-VE, Rev P
A calibration cannot be performed if there is a Fail alarm active except
in the case of when the alarm is triggered by low flow. Calibration can
start or continue during Low Sample Flow alarm status.
The span gas is the high calibration gas. The zero gas is the low cali-
bration gas.
Span
2
calibration, the order would be as follows:
2
To cancel a calibration at any time, select Abort and press ENTER/
MENU.
All, Oxygen, Combustibles, Methane
Define a time period for the system to recover from
reading calibration gas values to reading process
gases.
Enter the calibration gas values to match the span and
zero calibration gas cylinders you will use to calibrate
or verify your system.
, combustibles, methane)
2
Controller/Interface
|
2-17

Hide quick links: