Flow Calibration

Figure 2-17.1.
Flow Calibration should only
be done on a new sensor.
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
2. Press the Left or Right arrows to locate the Flow Calibration menu and
3. Press ENTER/MENU to continue with the calibration or press CANCEL to
4. Turn off the aspirator and calibration gas flow. Press ENTER/MENU to con-
5. The Flow Sensor Zero Calibration screen displays. Press ENTER/MENU to
6. The analyzer counts down to zero and when the zero calibration has been
7. You are prompted to inject the Cal gas. Be sure to use aspirator air or zero
2-26 | Thermox
®
AMEVision Display User Interface
The analyzer must have a flow sensor to configure flow calibration.
Flow sensor calibrations should only be performed on new flow sen-
sors or when the electronics have been changed. Use the same gas
used for O
span when calibrating the flow sensor.
2
Make sure that Flow is checked on the Select Sensor Options sub-
menu under Sensor Configuration.
tion Menu.
press ENTER/MENU.
exit the screen.
tinue or the CANCEL to abort the operation.
start the zero calibration.
completed, the analyzer will display the Flow Sensor Span Calibration
screen.
gas. Press ENTER/MENU to continue and access the Flow Sensor Span
Calibration screen.
PN 9000-165-VE, Rev P