Auto Cal Using the Calibration Start Menu

1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
2. Press the Left or Right arrows to locate the Start Calibration menu and
3. Navigate to the drop down list and press ENTER/MENU.
Figure 2-16.
Select calibration type.
4. Press the Right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Cal Gas Duration in seconds.
7. Enter the time for the Recovery Duration in minutes.
8. Press the Right arrow to continue.
9. When you are satisfied with your values, press ENTER/MENU
PN 9000-165-VE, Rev P
tion Menu.
press ENTER/MENU.
Select "Automatic. "
This is the Calibration Gas Values screen. If you have already entered
your calibration gas values in the Cal Gas Menu, they will be visible in
the fields. Any changes that you make to these values will be reflected
in the Cal Gas Menu as well.
to start the automatic calibration. The screen displays the Gas Value,
Cal Gas and Time Remaining. To abort the calibration, navigate to the
ABORT button and press ENTER/MENU.
If O
is selected for calibration, it will be the first gas calibrated and
2
the zero gas will be calibrated next, then combustibles and methane.
If combustibles is the first gas selected for calibration, zero gas will be
calibrated first, then combustibles and methane.
|