Control Diagnostics

1. Press the Left or Right arrows to locate the Control Diagnostics menu and
2. There are three screens that make up the Control Diagnostics menu:
Figure 3-1.
Set Analog Value screen
(Diagnostics Mode).
3-4 | Thermox
®
AMEVision Display User Interface
Diagnostics Mode is turned on when changes are made on this
menu so that the system does not accidentally trigger an auto calibra-
tion while troubleshooting.
press ENTER/MENU.
When you press ENTER/MENU key, the message "Diagnostics Mode:
Off" appears on the Status Bar at the bottom of the screen. This
changes to "Diagnostics Mode: On" if you make any changes. Press
CANCEL to exit and turn off Diagnostics Mode.
Analog I/O, Relays, and Close Contact.
This screen displays the three Outputs available, the Set Values, and the
Actual Values. Use the arrows to access the Set Value fields and the nu-
meric keypad to change the values.
Making a change to any of these fields will turn on the Diagnostics
Mode. Press CANCEL to exit Diagnostics Mode.
PN 9000-165-VE, Rev P