10. The screens will advance upon completion of each calibration to the next
11. When all gases have calibrated, the screen displays the Calibration Results

Perform a Remote Calibration

Using the Remote Cal Unit (RCU), you can manually accept the gas values
when each gas is stable by pressing the ENTER/MENU key, but the RCU
switches the gases on/off automatically.
1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
2. Press the Left or Right arrows to locate the Start Calibration menu and
3. Navigate to the drop down box, press ENTER/MENU key, and choose
4. Press the Right arrow to continue.
5. Select the gas(es) that you want to calibrate.
6. Enter the time for the Recovery Duration in minutes.
7. Press the Right arrow to continue.
8. When you are satisfied with your values, press ENTER/MENU to start the
2-24 | Thermox
®
AMEVision Display User Interface
gas to be calibrated.
that include % drift for gases calibrated as well as recovery time. If the
calibration is successful, a large GREEN check mark will be on the screen. If
it is not, a large RED 'X' is displayed.
The functions for remote calibration will only be visible if you have and
RCU and you have configured your sensor for Automatic Calibration
in the Sensor Configuration menu.
tion Menu.
press ENTER/MENU.
Remote.
This is the Calibration Gas Values screen. If you have already entered
your calibration gas values in the Cal Gas Menu, they will be visible in
the fields. Any changes that you make to these values will be reflected
in the Cal Gas Menu as well.
remote calibration. The screen displays the Gas Value, Cal Gas, and Status
(Ready/Wait). To abort the calibration, select the ABORT button and press
ENTER/MENU.
PN 9000-165-VE, Rev P