Calibrate the Analyzer

Perform a Manual Calibration

1. From the Main Menu screen, press ENTER/MENU to access the Calibra-
2. Press the Left or Right arrows to locate the Start Calibration menu and
3. You are prompted to select the gases to be calibrated. Select the individual
4. Enter the Recovery Duration time.
5. Press the Right arrow to advance to the next screen.
6. Enter the calibration gas values for the gases you have selected, press the
2-18 | Thermox
®
AMEVision Display User Interface
All measurements require a two-point calibration: span and zero, so
the zero will always be performed.
Calibrations are only allowed when the sensor is in Normal mode.
Check the Status Bar at the bottom of the HOME screen if you are
unsure of the sensor's status.
If you do not select a gas to calibrate, a warning displays indicating
that no gas was selected. Press CANCEL to exit the screen.
After a calibration is completed and if you have replaced the analyzer
cell since the last calibration, you will see a yellow screen indicating that
a cell change has been detected and asking if you would like to reset the
Cell Age. If you would, navigate to "Yes" and press ENTER/MENU. If you
would not, navigate to "No" and press ENTER/MENU.
tion Menu.
press ENTER/MENU.
gases or choose ALL.
Up and Down arrows to advance to the fields.
As a default, any gas values that you entered in the Cal Gas Values
menu will automatically be entered on this screen. You can edit these
values and they will be changed in the Cal Gas Values menu as well.
PN 9000-165-VE, Rev P