Energy Channel Recalibration Procedure (Mca) - Ametek Spectro MIDEX Original Operating Instructions

M-xrf spectrometer
Table of Contents

Advertisement

6.2.2

Energy channel recalibration procedure (MCA)

In order to recalibrate the instrument, proceed as follows:
1.
Ensure that the recalibration aid has been installed (see Section 6.2.1).
2.
In the software program, select the "Recalibration" menu item beneath the
green "Start" button and then select "MCA Recalibration".
The corresponding program window is displayed.
3.
Click on "OK" followed by "START".
The energy channel recalibration procedure is carried out automatically.
After the measurement is completed, a result window indicates that the
MCA recalibration procedure has been successfully completed.
The instrument is now calibrated.
SPECTRO MIDEX — 05.10.2017
Maintenance by the operator
— 27