5.4

Switching off the instrument

If the instrument is switched on and off very frequently, its service life will be
reduced (the X-ray tube being particularly affected).
Always leave the instrument switched on if possible.
Information
If the spectrometer is not used for longer than a day (weekends, vacation
shutdown), switch off the instrument. Whenever the instrument is switched on
again, the X-ray tube must be run under standby conditions for at least 20
minutes. Start measuring only after this period.
If you do not work with the system for an extended period of time, the X-ray
tube must be operated in standby mode for at least one hour once a month.
This procedure is also required if the instrument is not to be used for
measurements afterwards.
To switch off the instrument, proceed as follows:
1.
Click on the
software program.
A window opens.
2.
Click on "Shutdown". The connection between the PC and the instrument is
canceled.
3.
Confirm the security query and also confirm the shutoff of the high-voltage
supply.
SPECTRO MIDEX — 05.10.2017