C O M M A N D R E F E R E N C E
4
C H A P T E R
GPIB Command Reference
*OPC?
Operation Complete Query
Action
Response
Example
*PRE?
Action
Response
Bit Reference
0− Event Summary
1− Condition Summary
2− N/A
3− N/A
64
!
MPS-8033
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Places an ASCII character 1 into the intrument's Output Queue when all pending operations have
been finished.
1 − when all overlapped commands are complete.
*OPC?
Sets the parallel poll register enable bits for the format required by the 8033 in order that it may
participate in parallel polling from the controller.
The value must be between 0 and 65535. The Parallel Poll Enable Register is logically ANDed with
the Status Byte Register. If the result is non−zero the 'ist' (individual status) message is set true.
15 14131211 10
&
&
&
&
&
&
Logical
OR
15
14
13
12
11
10
IST TRUE
4− Message Available
5− Event Status Summary
6− Master Status Summary
7− Error Message Available
Copy of Status Byte
9 8
7 6 5 4 3 2 1 0
&
&
&
&
&
&
&
&
&
&
8
9
7 6 5 4 3 2 1 0
Parallel Poll Enable Register
8− N/A
9− N/A
10− N/A
11− N/A
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
12− N/A
13− N/A
14− N/A
15− N/A