L O C A L O P E R A T I O N
2
C H A P T E R
Introduction to the MPS-8033 Front Panel
MODE Section
When the CW switch is pressed the CW LED illuminates indicating the internal modulation function
is disabled and a continuous wave output is enabled.
The first execution of the INT MODULATION switch will return the modulation to the last selected
modulation frequency. Subsequent executions of the INT MODULATION switch will sequentially
select between modulation frequencies.
The MODE switches select between either continuous wave or internally modulated optical outputs.
The internally modulated output is a square wave pulse with a 50% duty cycle at one of three
frequencies; 270, 1000, and 2000 Hertz.
MODE SELECT
The CW switch is used to select Continuous Wave operation. When CW mode is selected, the
corresponding LED becomes lit and all modulation is disabled.
To switch from CW to pulse modulation, the INT MODULATION switch must be pressed. The first
execution of the INT MODULATION switch will return modulation to the last selected frequency.
Subsequent executions of the INT MODULATION switch will sequentially cycle through the
available modulation frequencies.
TTL Output
The rear panel TTL output is an electrical equivalent to the front panel modulated optical output.
OUTPUT Function
When the ON switch is pressed, the output on indicator flashes for two seconds before the laser is
enabled. Following the delay, the output on indicator fully illuminates and the display switches to the
actual measured output power.
10
!
MPS-8033
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Continuous
Wave Output
LED
Internal Pulse
Modulation
Frequencies
Figure 2.2 Mode Section
CW Switch
INT Modulation
Switch