ENABle:COND
Parameters
Notes
Examples
ENABle:COND?
Parameters
Response
Notes
Examples
7/01
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The ENABle:COND command sets the condition status enable register for the 8033. This register is
a filter which is logically ANDed to the condition register. If the result is non−zero, bit #1 in the Status
Byte is set to one.
An <nrf value> whose sum represents the enabled bits below:
1
Output is ON
2
Interlock is disabled
4096
Calculation Error
The condition enable register can be read by using the ENABle:COND? query.
The condition status can be monitored by the COND? query. If any of the enabled conditions are
true, bit #1 of the status register will be set.
The enable registers retain their values at power−up unless the power−on status clear flag is set true
(See *PSC in Chapter 3).
"ENAB:COND 3" action: allows the output on and interlock broken conditions to set bit #1 in the
Status Byte.
The ENABle:COND? query returns the value of the 8033 status condition enable register.
None.
The response data is the binary sum of the following:
1
Output is ON
2
Interlock is disabled
4096
Calculation Error
The condition enable register can be set by using the ENABle:COND command.
The condition status can be monitored by the COND? command.
"ENAB:COND?" response: 16385, means the output on and software error conditions are allowed
to update bit #1 of the Status Byte.
C O M M A N D R E F E R E N C E
GPIB Command Reference
C
D
F
8192
Hardware Error
16384
Software Error
32768
Software Checksum Error
C
D
F
8192
Hardware Error
16384
Software Error
32768
Software Checksum Error