ILX Lightwave MPS-8033 Series User Manual page 61

Precision fiber optic sources
Table of Contents

Advertisement

C O M M A N D R E F E R E N C E
4
C H A P T E R
GPIB Command Reference
CAL:USER1
Parameters
Notes
Examples
CAL:USER2
Parameters
Notes
Examples
CAL:USER3
SYNTAX DIAGRAM
Parameters
Notes
Examples
*CLS
44
!
MPS-8033
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The CAL:USER1 command will cause the 8033 to set the first calibration point into memory.
None.
In local control, this command is similar to pressing the SET REFERENCE switch while in the
calibration mode.
This command will cause the 8033 to set the first calibration point into memory. This command
must follow the CAL:START command and should not be used until the desired power level is
achieved. The 8033 will then await the CAL:USER2 command.
The proper test equipment should be set up in order for these commands to work. Calibration is
discussed in Chapter 6.
"CAL:USER1"
The CAL:USER2 command will cause the 8033 to set the second calibration point into memory.
None.
In local control, this command is similar to pressing the SET REFERENCE switch while in the
calibration mode.
The "CAL:USER2" command will cause the 8033 to set the second calibration point into memory.
This command must follow the CAL:START and CAL:USER1 commands and should not be used
until the desired power level is achieved.
The proper test equipment should be set up in order for these commands to work. Calibration is
discussed in Chapter 6.
"CAL:USER2"
The CAL:USER3 command will cause the 8033 to set the third calibration point into memory.
None.
In local control, this command is similar to pressing the SET REFERENCE switch while in the
calibration mode.
The "CAL:USER3" command will cause the 8033 to set the third calibration point into memory. This
command follows the CAL:USER2 command and should not be used until the desired power level is
achieved. The 8033 will then complete the calibration procedure and return the device to the state
before the calibration was initiated.
The proper test equipment is required. Calibration is discussed in Chapter 6.
"CAL:USER3"
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL