R E M O T E O P E R A T I O N
3
C H A P T E R
Command Syntax

Terminators

A program message terminator identifies the end of a command string. These are the valid
terminator sequences:
• <NL>
• <^END>
• <NL><^END>
Many computers terminate with <CR><NL><^END> (Carriage Return − New Line − EOI). A carriage
return (<CR>)is read as white space.
The 8033 terminates its responses with <CR><NL><^END>, unless you use the TERM command to
change it.
If you encounter problems with GPIB communications, the terminator string can sometimes be the
cause. Refer also to your GPIB interface (controller) manual.

Command Separators

You can put more than one command on the same line (same command string) if you separate them
with a semicolon. The semicolon can be preceded by one or more spaces. Examples:
MODE:DBM; LASER:POW -3.0 DBM
:DIS ON; :las:mod:freq?

Parameters

Some commands require a parameter. The parameter must be separated from the command by at
least one space.
The syntax symbol <nrf value> refers to the flexible numeric representation defined by the GPIB
standard. It means that you can represent numbers in integer or floating point form, or in
engineering/scientific notation. The IEEE−488.2 standard uses the names NR1, NR2, and NR3
respectively to denote "integer", "floating point", and "scientific notation". For example the number
"twenty" may be represented by any of the following ASCII strings:
Integer
Floating point
Scientific notation
22
!
MPS-8033
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
20
+20
20.0
+20.0
2.0E+1
+2.0E+1
2.0e+1
+2.0e+1
NR1
NR2
NR3