ILX Lightwave MPS-8033 Series User Manual page 62

Precision fiber optic sources
Table of Contents

Advertisement

Action
Note
Example
COND?
Parameters
Response
1
2
4096
Notes
Examples
DISplay
Parameters
Notes
Examples
7/01
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Clears status event registers: Event Status, Event Status Enable, and Error Queue.
Useful to clear registers before enabling service requests (SRQ).
*CLS
The COND? query returns the value of the 8033 status condition register.
None.
The response data is the binary sum of the following:
Output is ON
8192
Interlock is disabled
16384
Calculation Error
32768
The condition status may be constantly changing. This register is not cleared when queried or when
*CLS is issued.
The conditions which are reported to the status byte are filtered by the condition enable register
(See ENAB:COND).
"COND?"
response: 1, means the laser output is ON.
The DISplay command enables or disables (turns off) the front panel display and LED indicators.
The <nrf value> of 0 or OFF to disable the display, 1 or ON to enable the display.
When the display is enabled, the value displayed will depend upon the present mode of operation.
When the display is disabled, the display contains a blank message and all of the front panel LEDs
will be turned off. This command is useful when it is necessary to work in a dark environment.
"DIS ON"
action: display is turned on.
"DIS 0"
action: display is turned off.
C O M M A N D R E F E R E N C E
GPIB Command Reference
C
D
F
Hardware Error
Software Error
Software Checksum Error
C
D
F
4
C H A P T E R
OMMON
D
EVICE
EPENDENT
P
RONT
ANEL
OMMON
D
EVICE
EPENDENT
P
RONT
ANEL
45
MPS-8033
!

Advertisement

Table of Contents
loading

Table of Contents