ROBOTICS VERO
2.1.4
Robotics with iBeam evo Couch Top
Safety Notes
The maximum distributed load permitted on Robotics combined with the iBeam evo Couch
Top is 135kg.
The maximum point load at the end of the cranial end of theiBeam evo Couch Top is 100kg.
The maximum load on the iBeam evo extension is 70kg.
Do not step on the iBeam evo couch top.
Warning Labels
The following warning labels are attached on Robotics Vero or its sub-components:
Label
The following warning label is attached to the iBeam evo Couch Top:
Label
Hardware User Guide Rev. 1.1 Robotics Vero Ver. 2.1
Explanation
• The maximum point load at the
end of the cranial end of thei-
Beam evo Couch Top is 100kg.
• The maximum load on the iBeam
evo extension is 70kg.
• Do not step on the Imaging
Couch Top.
The maximum distributed load per-
mitted on Robotics combined with
the iBeam evo Couch Top is 135kg.
Explanation
The patient's weight must not ex-
ceed 135 kg, evenly distributed
over the iBeam evo Couch Top.
21