Robotics Hardware and Software Components
2.1.3

Robotics with Imaging Couch Top

Safety Notes
The maximum distributed load permitted on Robotics Vero combined with the Imaging
Couch Top (with extension) is 135 kg.
The maximum point load at the cranial end of the Imaging Couch Top (no extension) is
92,5kg.
The maximum load on the Extension of the Imaging Couch Top is 40kg.
Do not step on the Imaging Couch top.
Warning Labels
The following warning labels are attached on Robotics Vero or its sub-components:
Label
The following warning label is attached to the Imaging Couch Top:
Label
20
Explanation
• The maximum allowable point
load applied to the cranial end of
the Imaging Couch top (without
frameless extension) is 92.5kg.
• The maximum allowable load for
the Imaging Couch top frame-
less extension is 40 kg.
• Do not step on the Imaging
Couch top.
The patient's weight must not ex-
ceed 135 kg, evenly distributed
over the Imaging Couch top
Explanation
The patient's weight must not ex-
ceed 135 kg, evenly distributed
over the Imaging Couch Top.