EMC Notes

4.2
EMC Notes
Safety Notes
NOTE: This Medical Electrical Equipment requires special precautions regarding EMC and must
be installed and put into service by Brainlab service staff and/or authorized service partners.
NOTE: Portable and mobile RF communications equipment can affect Medical Electrical
Equipment.
The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by brainlab as replacement parts for internal
components, may result in increased emissions or decreased immunity of the equipment
or system.
Do not modify system setup or configuration of system components! in case of
modifications concerning electrical installation adjacent to the equipment, e.g. routing of
cables, verify that the equipment operates as intended.
The device may be subject to electrical interference by other equipment, even if that other
equipment complies with CISPR emission requirements.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Robotics Vero is intended for use in the electromagnetic environment specified below. The
customer or the user of the Robotics Vero should assure that it is used in such an environment.,
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flick-
er emissions
IEC 61000-3-3
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and
Robotics Vero
Rated Maximum Output
Power of Transmitter
W
50
Compliance
Electromagnetic environment - guidance
Robotics Vero uses RF energy only for its internal
function. Therefore, its RF emissions are very low
Group 1
and are not likely to cause any interference in near-
by electronic equipment.
Robotics Vero is suitable for use in all establish-
Class A
ments other than domestic, and may be used in do-
mestic establishments and those directly connected
to the public low-voltage power supply network that
A
supplies buildings used for domestic purposes, pro-
vided the following warning is heeded:
Warning: This equipment/system is intended for use
by health care professionals only. This equipment/
system may cause radio interference or may disrupt
Complies
the operation of nearby equipment. It may be neces-
sary to take mitigation measures, such as re-orient-
ing or relocating Robotics Vero or shielding the lo-
cation.
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
d
=
1.17 P
d
Hardware User Guide Rev. 1.1 Robotics Vero Ver. 2.1
800 MHz to 2,5 GHz
=
1.17 P
d
=
2.34 P