Disposal of the medical device

At the end of its service life:
 Consult the relevant waste disposal company for
appropriate disposal.
 Observe the applicable laws and regulations.
18

Fault-Cause-Remedy

Fault
Cause
No mist (flow-
No drive gas
meter or toggle
switch valve)
Leakage in
nebulizer ele-
ment
Blocked atom-
izer
Leakage in at-
omizer
Droplets too
Pressure of
large (flowme-
drive gas too
ter)
low
Leakage in at-
omizer
Disinfect and clean device before maintenance or re-
pair.

Technical data

Operating principle
Remedy
Drive gas
Open metering
valve
Check gas
Gas consumption
supply
Flowmeter
Screw medica-
Toggle switch valve
tion container
on tightly
Nebulizer output
Replace atom-
at 6 L/min
izer
at 12 L/min
Check O-ring
Connecting thread
on connector,
Operating temperature
replace, if nec-
essary
Storage temperature
Open metering
Relative humidity
valve further
Sound pressure level
Weight
Check O-ring
Nebulizer element
on connector,
with mask
replace, if nec-
essary
Toggle switch valve
Materials used
Mask and housing
Atomizer and medica-
tion container
Classification
as per EC Directive
93/42/EEC Annex IX
UMDNS-Code
Universal Medical Device
Nomenclature System –
Nomenclature for medical
devices
Venturi nebulizer
O
or medical com-
2
pressed air (Air)
1 to 16 L/min
10 L/min
>2 mL/10 min
>3 mL/10 min
M 15 x 1
10 °C to 40 °C
–20 °C to 70 °C
0 to 90 %
60 dB (A) at 15 L/min
about 110 g
about 350 g
polyamide
polysulphone
Class II a
10-046
Instructions for use Inhalette