The power supply can accept values 5% higher than the values Tables
7-3 and 7-4, however it is not recommended to program the power
Rated
Output Voltage
(V)
8
10
16
20
30
40
60
80
100
150
300
600
Table 7-5: OVP Programming Range
7.7.7
Status Control Commands
Refer to Section 7-8 for definition of registers.
#
Command
1
STT?
Reads the complete power supply status.
Returns ASCII characters representing the following data, separated by commas:
MV<actual (measured) voltage>
PV<programmed (set) voltage>
MC<actual (measured) current>
Example response: MV(45.201), PV(45), MC(4.3257), PC(10), SR(30), FR(00)
2
FLT?
Reads Fault Conditional Register. Returns 2-digit hexadecimal (hex)
3
FENA
Set Fault Enable Register using 2-digit hexadecimal (hex)
4
FENA?
Reads Fault Enable Register. Returns 2-digit hexadecimal (hex)
5
FEVE?
Reads Fault Event Register. Returns 2-digit hex. Clears bits of Fault Event Register.
6
STAT?
Reads Status Conditional Register. Returns 2-digit hexadecimal (hex)
7
SENA
Sets Status Enable Register using 2-digit hexadecimal (hex)
8
SENA?
Reads Status Enable Register. Returns 2-digit hexadecimal (hex)
9
SEVE?
Reads Status Event register. Returns 2-digit hex. Clears bits of Status Event register.
supply beyond its rated values.
OVP
OVP
Minimum
Maximum
(V)
(V)
0.5
10.0
0.5
12.0
1.0
19.0
1.0
24.0
2.0
36.0
2.0
44.0
5.0
66.0
5.0
88.0
5.0
110
5.0
165
5.0
330
5.0
660
83-517-000 Rev. A
NOTE
Rated
Output Voltage
(V)
8
10
16
20
30
40
60
80
100
150
300
600
Table 7-6: UVL Programming Range
Description
PC<programmed (set) current>
SR<status register, 2-digit hex>
FR<fault register, 2-digit hex>
66
UVL
UVL
Minimum
Maximum
(V)
(V)
0
7.60
0
9.50
0
15.2
0
19.0
0
28.5
0
38.0
0
57.0
0
76.0
0
95.0
0
142
0
285
0
570