Global Output Commands - TDK-Lambda GENESYS GEN 2400W Series Technical Manual

1u programmable dc power supplies
Hide thumbs Also See for GENESYS GEN 2400W Series:
Table of Contents

Advertisement

NOTES:
1. In "Advanced Parallel" mode (refer to Section 5.15.2), "n" is the total system Output current.
2. In "Advanced Parallel" mode, "MC?" returns the Master unit Output current multiplied by the "Number of
Slave units + 1".
7.7.6

Global Output Commands

7.7.6.1
General
All power supplies, even if not the currently addressed power supply, receiving a Global command will execute
that Global command. No response (to the controller/PC issuing the command) will be returned to the
controller/PC. The controller/PC issuing the Global command will be responsible to delay any other
communications until the command is executed. The suggested delay is 200 milliseconds minimum.
If the Global command contains an error, "out of range values" for example, no error report will be sent to the
issuing controller/PC.
#
Command
1
GRST
2
GPV n
3
GPC n
4
GOUT
5
GSAV
6
GRCL
Model
Minimum
VOrated
8
0.0000
10
00.000
16
00.000
20
00.000
30
00.000
40
00.000
60
00.000
80
00.000
100
000.00
150
000.00
300
000.00
600
000.00
Table 7-3: Voltage programming range
Global Reset. Brings the power supply to a safe and known state:
Output voltage: 0V, Output current: 0A, OUT: Off, Local/Remote: Remote, AST:
OFF,
OVP: Maximum, UVL: 0V.
The conditional register (FLT and STAT) are updated. Other registers are not
changed.
Non-latching faults (FB, OVP, SO) are cleared, OUT fault stays
Sets the Output voltage value in Volts. The range of Output voltage values is shown in
Table 7-3. 'n' may be up to 12 characters plus decimal point
Program the Output current value in Amperes. The range of Output current values is
shown in Table 7-4. 'n' may be up to 12 characters plus decimal point
Turns the output to ON or OFF:
"OUT 1 (or OUT ON)" = Turn Output ON
"OUT 0 (or OUT OFF)" = Turn Output OFF; clears CV and CC bits in the Status
Condition (STAT) register.
OUT ON will respond with "E07" if the output cannot be turned ON because of a
latching fault (OTP, AC, ENA, SO) shutdown.
Saves "Present Settings". Same settings as power-down last settings listed in Section
5.13 except the unit Address and Baud rate are not saved to EPROM memory. These
settings are erased when the power supply AC Input power is switched OFF and the
new 'Last Settings" are saved.
Recalls "Last Settings". Settings are from last AC Input power-down or from last 'SAV'
or
'GSAV' command. Unit Address and Baud rate are not recalled so serial
communication is not interrupted.
Maximum
(V)
(V)
8.000
10.000
16.000
20.000
30.000
40.000
60.000
80.000
100.00
150.00
300.00
600.00
83-517-000 Rev. A
Description
Model
Minimum
IOrated
300
240
150
120
80
60
40
30
24
16
8
4
Table 7-4: Current programming range
65
Maximum
(A)
(A)
000.00
300.00
000.00
240.00
000.00
150.00
000.00
120.00
00.000
80.000
00.000
60.000
00.000
40.000
00.000
30.000
00.000
24.000
00.000
16.000
0.0000
8.0000
0.0000
4.0000