12 Ethernet connection specification
Measurement results
The measurement result query gives the measured and calculated measurement val-
ues from the chosen sensor.
Request ID
0x00000004
Request data
0x00000000
Request data
0x00000001
Response keys
Status
Slope
PTraw
LED
RHsens
nD
CONC
Tsens
T
Traw
CCD
CALC
DTR status
Request ID
0x00000006
Request data
(none)
Response key
*Volt1
*Volt2
*DTRtemp
Out1uA
Out2uA
Switches

12.5.3 Error message specification

If the server (DTR) does not recognize the request or cannot fulfill it, it responds with
an error message. The error message has the following keys:
: integer, error code
*Error
: integer, error code
*Error
: integer, error code
*Error
: string, error details
ErrorMsg
There may also be error-dependent extra keys. Other error codes may be returned.
is to be handled as unknown request. Codes with higher numbers refer
0x00000003
to internal errors; contact Vaisala for more information on these.
: sensor A
: sensor B
: string, sensor status message
: float, image quality factor (QF)
: integer, PT1000 value
: float, sensor led value
: float, sensor internal humidity
: float, calculated n
value
D
: float, final concentration value
: float, sensor internal temperature
: float, process temperature (with temperature bias)
: float, process temperature (without bias)
: float, image shadow edge
: float, calculated concentration value
: float, DTR internal voltage 1
: float, DTR internal voltage 2
: float, DTR internal temperature
: integer, mA output 1 in uA
: integer, mA output 2 in uA
: hex string (e.g. "0x00"), switch status as a bit field
: Unknown request
0x00000000
: Invalid request (request recognized,
0x00000001
invalid request data)
: No sensor (sensor(s) not connected
0x00000002
to DTR)
207