Prism Wash Nozzles - Vaisala K-PATENTS PR-23 Series Instruction Manual

Process refractometer
Hide thumbs Also See for K-PATENTS PR-23 Series:
Table of Contents

Advertisement

26

4.2.3 Prism wash nozzles

When selecting a wash nozzle for a compact refractometer, take into account both
the wash medium and the flowcell model: flowcells with larger pipe diameters need
longer wash nozzles. Figure 4.9 below shows a wash nozzle for a flowcell and gives
the measurements and part numbers for each nozzle type.
Flowcell -H10 or -H15
A
steam
64.75
water
75
pressurized water
75
Flowcell -H20 or -H25
A
Mark
steam
72.15
C
water
97
pressurized water
97
Flowcell -H30
A
steam
103
water
113
pressurized water
113
Flowcell -H40
A
steam
133
water
143
pressurized water
143
Figure 4.9 Wash nozzles for flowcell AFC-HSS-XXX-XX-NC
Figure 4.10 shows how the nozzle is mounted in a flow cell (-NC with stud for a wash
nozzle). Note: See Section 9.3.5 for more information on flowcells.
For probe refractometers, select the wash nozzle according to wash medium and
refractometer model, see Table 4.1 below.
Figure 4.11 shows the mounting of the wash nozzle for Sanitary probe refractometer
PR-23-AP. Figure 4.12 shows the mounting of the wash nozzle for Process refractome-
ter PR-23-GP.
Part no.
General tolerances
SFS 4011 FINE
B
part nr
4.0
PR-3365
2.5
PR-3369
1.5
PR-3368
B
part nr
Rev.
4.0
PR-3375
MRP nrot muuttuneet
2.5
PR-3379
1.5
PR-3378
B
part nr
4.0
PR-3393
2.5
PR-3394
1.5
PR-3395
B
part nr
4.0
PR-3390
2.5
PR-3391
1.5
PR-3392
Drawing no.
Title/Name, designation
Scale
Join
1 1
:
PR-23 instruction manual
A
Standard
File
756.drw
Drawing description
AFC nozzles
Drawing number
Date
Rev. by
Appr.
0.8
Quantity
Dimensions, model no.
Material
AISI316L
30.8.2004 HS
Drawn

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents