Configuration And Calibration; Configuring Output Signal Damping; Exponential Damping - Vaisala K-PATENTS PR-23 Series Instruction Manual

Process refractometer
Hide thumbs Also See for K-PATENTS PR-23 Series:
Table of Contents

Advertisement

6 Configuration and calibration

6 Configuration and calibration
All changes of configuration and calibration are made through the Calibration menu
selected from the Main menu by
Password: It may be necessary to enter a password before proceeding to the Calibra-
tion menu. The password is printed on the title page of this manual. The password
function is activated and deactivated via the
menu.
By default the password is activated.

6.1 Configuring output signal damping

The Outputs display also provides the possibility to enter signal damping to diminish
the influence of process noise. The damping is applied to the CONC value (and thus
the output signal) of the current sensor (see upper edge of the display to check which
sensor is currently chosen and switch in the Outputs display if necessary).
The PR-23 offers three types of signal damping. The damping parameter is set sep-
arately through the Outputs menu selected from the Calibration menu by
What the damping time means in practice depends on the damping type.

6.1.1 Exponential damping

Exponential (standard) damping works for most processes and is the standard choice
for slow and continuous processes. The factory setting is always exponential damping,
access the
3 DAMPING TYPE
In the exponential damping (standard damping), the damping time is the time it takes
for the concentration measurement to reach half of its final value at a step change. For
example, if the concentration changes from 50 % to 60 % and damping time is 10 s, it
takes 10 seconds for the DTR to display concentration 55 %. A damping time of 5–15
seconds seems to work best in most cases, the factory setting is 5 seconds. Use the
menu item to set the damping time. Figure 6.1 shows how exponential
DAMPING TIME
damping time affects the measurement.
13.0
12.9
12.8
12.7
12.6
12.5
12.4
12.3
12.2
12.1
12.0
0
10
.
5 CALIBRATION
6 PASSWORD
command to switch between different damping algorithms.
5 s
10 s
20 s
20
30
40
50
60
70
Time [s]
Figure 6.1 Exponential damping
command in the Calibration
2 OUTPUTS
80
90 100 110 120
41
.
4

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents