6 Configuration and calibration
Figure 6.9 Arrival to Switch 1's sensor selection menu, sensor A selected
3
WASH STOP
4
REMOTE WASH
5
SCALE SELECT
6
CALIBRATION SEAL
7. If you chose Scale select as switch function, return to the Switch menu (if not re-
turned automatically) and choose
the chemical curve assigned to the switch. See Section 6.4.1 for more information
on chemical curves and chemical curve parameters.
8. If necessary, the chemical curve assigned to a switch can be adjusted by field cali-
bration parameters. Select
ters.See Section 6.4.3 for more information on field calibration and field calibra-
tion parameters.

6.3.3 Configuring mA outputs

For the electrical properties of the two output signals, see Section 3.3.3, "Connecting
the Indicating transmitter".
process stops (as indicated by contact closure).The wash is
repeated when the process restarts (if the stop lasts over
60 seconds). The signal is on hold between washes.When
used with an external independent timer, contact closure
holds the output signal.
Switch closure prevents wash cycle. It can be used to pre-
vent wash action when the process pipe is empty. The mes-
sage
will appear when a wash cycle is initiated.
WASH STOP
At switch closure the system waits for an external wash
command before initiating wash.
Any chemical curve and associated field calibration scale
can be selected by switch closure. The scales assigned to
each switch independently.
Contact closure prevents access to calibration and config-
uration ("external password"). Can be used to seal the cal-
ibration.
3 SCALE CHEMICAL
in the Switch menu to enter the parame-
4 SCALE FIELD
to enter the parameters for
49