Linear Damping; Slew Rate Limit; Configuring Output Signal Hold Functionality - Vaisala K-PATENTS PR-23 Series Instruction Manual

Process refractometer
Hide thumbs Also See for K-PATENTS PR-23 Series:
Table of Contents

Advertisement

42

6.1.2 Linear damping

If the process has fast step changes, linear (fast) damping gives shorter settling time.
In the linear damping (fast damping), the output is the running average of the signal
during the damping time. After a step change the signal rises linearly and reaches the
final value after the damping time. The linear damping gives the best trade-off be-
tween random noise suppression and step change response time. Use the
menu item to set the damping time. Please note that for similar noise suppres-
TIME
sion a longer damping time has to be specified than for the exponential damping.
Figure 6.2 shows how linear damping time affects the measurement.
13.0
12.9
12.8
12.7
12.6
12.5
12.4
12.3
12.2
12.1
12.0
0
10

6.1.3 Slew rate limit

If the process signal has short erroneous high or low peaks, the slew rate limiting can
be used to cut their effects. The slew rate damping limits the maximum change for
the output signal in one second. It should be noted that the slew rate limit damping is
recommended for random noise suppression as it is non-linear.
The slew rate limit can be set through the menu item
pend on the concentration unit but are typically from 0.05 % to 1 % when the concen-
tration is measured in percentages. Figure 6.3 gives an example of different slew rate
limits.
Note: Avoid overdamping, the signal should not be made insensitive.

6.2 Configuring output signal hold functionality

The instrument can be configured to temporarily hold its measurement result in three
different cases.
7 s
15 s
30 s
20
30
40
50
60
70
Time [s]
Figure 6.2 Linear damping
PR-23 instruction manual
4 DAMPING
80
90 100 110 120
. Typical values de-
5 SLEW RATE