Motor Angle Adjustment; Reusable Bearing Hub Insertion For Open-Top Liners - Thermo Scientific HyPerforma DS 300 User Manual

Single-use mixer (s.u.m.)
Table of Contents

Advertisement

Chapter 2
|
Setup and operating information
Figure 2.25. Motor angle
adjustment
Release knob
for adjustable
motor head
Thermo Scientific
2. The correct height can be determined using the height indicator
labels on the vertical cable management housing. The height
should fall within the white area for the size of the drum in use.
Note that the labels assume the drum is on a dolly, and are not
accurate if the drum is not on a standard dolly.

2.6.3 Motor angle adjustment

Certain models are equipped with motors that can be adjusted to
different angles. To adjust the angle of these motors, follow the steps
below.
1. Hold the handle on the motor, and pull the knob on the side of
the motor block until the motor is free to rotate relative to the
horizontal plane.
2. There are pin holes every 5°, from 0–30°. Rotate the motor head,
and release the knob when the motor angle is correct. Ensure that
it locks into the desired position (Figure 2.25). CAUTION: Never
adjust the motor angle while the motor is running.

2.6.4 Reusable bearing hub insertion for open-top liners

A reusable bearing hub is needed when using open-top containers.
Use the following steps to insert the reusable hub.
1. Open the latching mechanism under the motor head and slide the
hub into the holder (Figure 2.26).
HyPerforma DS 300 Single‑Use Mixer User's Guide
|
31